361T B/W TRANSMISSION DENSITOMETER Operation Manual

Page 2

CHAPTER ONE 1-2 [6] READ HEAD - is the component that contains the optics which comes in contact with your film when taking measurements. [7] POWER SW

Page 3 - RoHS/WEEE

OVERVIEW AND SETUP 1-3 Features Electronic Filter Selection - eliminates the problems that mechanical filter wheels create. Large Liquid Crystal Displ

Page 4 - DENSITOMETER

CHAPTER ONE 1-4 Apply Power and Display Angle Adjustment NOTE: If the unit has been stored in an abnormal (cold) environment, DO NOT apply power to t

Page 5

361T DENSITOMETER 2-12. User Interface Key Description The Keyboard consists of three key switches, the [FUNCTION] d key, the [COLOR] c key, and the

Page 6

CHAPTER TWO 2-2 DISPLAY CONTRAST • Rotate the wheel to the back (+) to increase contrast or to the front (-) to decrease contrast. Function Selecti

Page 7 - Warranty Information

USER INTERFACE 2-3 NOTE: If DEN is the only function that can be activated, refer to "311 Emulation" to turn it Off. Color Selection The

Page 8

CHAPTER TWO 2-4 Reference and Base Entry Reference values are values, which are subtracted from each density measurement to display a density differen

Page 9 - Overview and Setup

USER INTERFACE 2-5 Base Entry The Base value is usually set to 0% for +DOT and 100% for -DOT, or adjusted to a value equal to a known minimum (base) d

Page 10 - CHAPTER ONE

CHAPTER TWO 2-6 MESSAGE REASON condition (if lamp was OK or replaced) by pressing [FUNCTION], then [COLOR], then [FUNCTION]. MEM TEST: The memory i

Page 11 - Unpack and Inspect

361T DENSITOMETER 3-13. Measurement Procedures Density Function The DEN function allows you to take density and density difference measurements.

Page 13 - User Interface

CHAPTER THREE 3-2 -DOT Function −Dot allows negative dot measurements with zeroing to the Base Dot value. −Dot can be calculated for a Base equal t

Page 14 - Displayed

MEASUREMENT PROCEDURES 3-3 1. Repeatedly depress the [FUNCTION] key until "SEQUENCE" is displayed. (If sequence does not appear, refer to &

Page 15 - Color Selection

CHAPTER THREE 3-4

Page 16 - Reference and Base Entry

361T DENSITOMETER 4-14. Calibration General Information The X-Rite 361T is designed for long life and extremely stable measurements. To follow good

Page 17 - Display Messages

CHAPTER FOUR 4-2 5. Enter the correct Cal Hi value for Ortho (as marked on the step tablet) by pressing and holding dc (ZERO) down and using the d (

Page 18 - 311 Emulation

INSTRUMENT CALIBRATION 4-3 N-Factor Adjustment Procedure NOTE: Transmission dot "N-factor" is calibrated at the factory and therefore sho

Page 19 - Measurement Procedures

CHAPTER FOUR 4-4 Quick CAL™ Procedure Quick Cal™ provides you with an easy means of re-establishing zero (Calibration Low). This method is included

Page 20 - -DOT Function

361T DENSITOMETER 5-15. Mode Selection General Information The Mode function controls four factors: x10 On/Off, I/O Port set-up, Sequence setup, and

Page 21 - MEASUREMENT PROCEDURES

CHAPTER FIVE 5-2 I/O Port Selection Your X-Rite 361T comes equipped with a serial port that allows data to be transmitted and/or received by the 361T

Page 22 - CHAPTER THREE

MODE SELECTION 5-3 The procedure for setting the I/O port options is shown below. 1. Depress [FUNCTION] and [COLOR] keys at the same time until "

Page 23 - Calibration

361T DENSITOMETER Federal Communications Commission Notice This equipment has been tested and found to comply with the limits for a Class A digital d

Page 24 - CHAPTER FOUR

CHAPTER FIVE 5-4 Setting the Sequence Structure Sequence allows you to setup a measurement routine that fits your individual quality control requireme

Page 25 - INSTRUMENT CALIBRATION

MODE SELECTION 5-5 The procedure for setting Sequence structure is shown below. 1. Depress [FUNCTION] and [COLOR] keys at the same time until "N

Page 26 - Quick CAL™ Procedure

CHAPTER FIVE 5-6 Read Lamp ON/OFF This feature allows the densitometer to automatically turn the Read lamp OFF after 2 hours of instrument non-use. Ha

Page 27 - Mode Selection

361T DENSITOMETER 6-16. Serial Interface Interconnect and Definition The 25-pin connector used for serial input/output is a DB25S type. Below is the

Page 28 - I/O Port Selection

CHAPTER SIX 6-2 A typical interconnection (in its simplest form) between the 361T and a computer is shown below. 361T COMPUTER W/DB25 Receive Data

Page 29 - MODE SELECTION

SERIAL INTERFACE 6-3 Instruction Format In general, the format used as the command string for serial input is as follows: 1 2 3 4 DATA — ADDRESS —

Page 30 - CHAPTER FIVE

CHAPTER SIX 6-4 J [KEY ENABLE] is used to enable or disable all keys (with the exception of the Read button) from altering the operation of the unit.

Page 31

SERIAL INTERFACE 6-5 L [LOAD] writes Data to the external RAM memory location Address. Data must be designated as 2 hexadecimal characters and the Ad

Page 32 - Read Lamp ON/OFF

CHAPTER SIX 6-6 T [BAUD] is used to change the baud rate of the 361T. The unit must receive the following as the address and the action codes portion

Page 33 - Serial Interface

SERIAL INTERFACE 6-7 W [WRITE] writes Data to the Address location of the microprocessor's internal RAM memory. This action causes the 361T to r

Page 34 - Serial Input Commands

361T DENSITOMETER ii CAUTION: To prevent electrical shock. DO NOT remove cover. No user serviceable parts inside. Refer servicing to qualified service

Page 35 - Instruction Format

CHAPTER SIX 6-8 Z [DISPLAY ENABLE] is used to enable or disable the display from being updated with data that occurs from normal operation. This swit

Page 36 - CHAPTER SIX

SERIAL INTERFACE 6-9 Internal RAM Data Addresses The Internal RAM Data Addresses are used to access parameters not handled by action codes. The follow

Page 37 - SERIAL INTERFACE

CHAPTER SIX 6-10 External RAM Data Addresses The External RAM Data Addresses are used to access data values, reference values, and other values. The f

Page 38

SERIAL INTERFACE 6-11 NAME ADDRESS DATA PARAMETERS Step 34,35 24A , 24B 1 Byte Step 36,27 24C , 24D Sequence Data ORTHO UV Step 0 250

Page 39

CHAPTER SIX 6-12 NAME ADDRESS DATA PARAMETERS CAL HI ORTHO 05A , 05B 2 Bytes UV 066 , 067 2 Bytes SKIP PROMPTS 700 1 Byte Wr

Page 40

361T DENSITOMETER 7-17. Maintenance General The X-Rite 361T is covered by a one-year limited warranty (excluding lamps) and should be referred to th

Page 41 - Internal RAM Data Addresses

CHAPTER SEVEN 7-2 Aperture Replacement & Cleaning APERTURE REPLACEMENT 1. Remove aperture [1] by wedging edge of aperture upwards with fingernail

Page 42 - External RAM Data Addresses

MAINTENANCE 7-3 Beam Splitter Cleaning 1. Remove aperture as described in "Exterior Cleaning". 2. Clean residue from Beam Splitter [1] wit

Page 43

CHAPTER SEVEN 7-4 Lamp Replacement AS A ROUTINE MAINTENANCE PROCEDURE, replace the read lamp after every 1200 hours of operation. This allows approxim

Page 44

MAINTENANCE 7-5 ALIGNMENT CHECK 8. Set unit upright, plug line cord into wall receptacle, and turn power switch to "ON" position. 9. Locat

Page 45 - Maintenance

361T DENSITOMETER iii Table of Contents 1. Overview and Setup... 1-1 Instrument

Page 46 - CHAPTER SEVEN

CHAPTER SEVEN 7-6 ALIGNMENT CHECK (continued) 12. Unplug line cord then loosen two Phillips-head screws 6 in bottom of lamp assembly. 13. Set front

Page 47 - Optics Cleaning

MAINTENANCE 7-7 Fuse Replacement [DISCONNECT LINE CORD BEFORE SERVICING] 1. Insert flatblade screwdriver into slot of fuse cap [1]. 2. Push in and t

Page 49 - MAINTENANCE

361T DENSITOMETER 8-18. Appendix Technical Specifications PARAMETER SPECIFICATION Display 8 character LCD, 9.4mm H w/ EL backlighting Measuring Rang

Page 50

CHAPTER EIGHT 8-2 PARAMETER SPECIFICATION Response Orthochromatic X-RITE Ortho Ultraviolet X-RITE U.V. Visual (optional) X-RITE Visual Color Tem

Page 51 - Fuse Replacement

APPENDIX 8-3 Optional Equipment There are six models of the 361 Reflection Head that are available to interface with the 361T Transmission unit. Model

Page 52

CHAPTER EIGHT 8-4 Interface Cables available: P/N 361-69 Transmission/Reflection Interface Cable P/N 361-83 DB25P to 10 Circuit Modular Interfac

Page 56

361T DENSITOMETER iv 5. Mode Selection... 5-1 General Information...

Page 57

Corporate Headquarters - USA 4300 44th Street SE Grand Rapids, Michigan 49512 Phone 1 800 248 9748 or 1 616 803 2100 Fax 1 800 292 4437 or 1 616 803

Page 58

361T DENSITOMETER v Proprietary Notice The information contained in this manual is derived from patent and proprietary data of X-Rite, Incorporated. T

Page 59

361T DENSITOMETER vi This agreement shall be interpreted in accordance with the laws of the State of Michigan and jurisdiction and venue shall lie wit

Page 60 - P/N 361T-500 Rev. JJ

361T DENSITOMETER 1-11. Overview and Setup Instrument Description The X-Rite 361T is a B/W Transmission densitometer designed to meet the quality co