Maintenance
............................
8.3
U
8.3 Replacement of measuring lamp and tube
Procedure:
Undo mounting screws and remove measuring aperture.
Manually set adjustment knob to position A.
Withdraw measuring module manually until adjustment knob is in
position B (matched arrows).
Undo screws and remove tube.
Replace lamp module (snap connection).
Insert tube and screw tight.
Install and then adjust measuring aperture (see chapter 8.2).