Vue de la page 92
5
Settings
............................
5.6.4 – 5.6.4.1
Press the <Measurement key> to get into the entry field for the color Black.
Turn the <Control ball> to correct the calibration value until it corresponds with the target value
indicated on the calibration card.
Terminate the entry by pressing the <Measurement key>.
Correct the calibration values of the colors Cyan, Magenta and Yellow one after the other.
The graphics bar then jumps to the menu line Accept data.
Store the calibration data by pressing the <Measurement key>.
The display shows Check calibration in reverse video.
• • • • •
5.6.4.1 Reject data
Procedure:
Select the menu line Reject data by turning the <Control ball> and confirm the action by pressing
the <Measurement key>.
The new data will not be stored; previously stored data are retained.
The menu lines Return, Check calibration and New calibration appear on the display.
Select the desired menu line and confirm the selection by pressing the <Measurement key>.