X-Rite D19C Standard > D19C Full Upgrade Manuel de service Page 80

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 164
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 79
Settings
............................
5.5.3.1 – 5.5.3.1.2
Select the menu line 3.Screen percent and confirm the selection by pressing the <Measurement
key>.
The display shows the entry field for the 3rd halftone reference value.
Enter the 3rd halftone reference value by turning the <Control ball> and confirm the entry by
pressing the <Measurement key>.
5.5.3.1.1 Accept data
Procedure:
Select the menu line Accept data by turning the <Control ball> and confirm the action by pressing
the <Measurement key>.
The menu line Return is shown in reverse video on the display.
Press the <Measurement key> several times to return to the main menu.
5.5.3.1.2 Reject data
Procedure:
Select the menu line Reject data by turning the <Control ball> and confirm the action by pressing
the <Measurement key>.
The new data will not be stored; previously stored data are retained.
The menu line Return is shown in reverse video on the display.
Press the <Measurement key> several times to return to the main menu.