X-Rite D19C Standard > D19C Full Upgrade Manuel de service Page 66

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 164
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 65
Settings
............................
5.3.1.2 – 5.3.1.3
5.3.1.2 Setting of default values
Procedure:
Select the menu line Set default values by turning the <Control ball> and press the <Measurement
key>.
The display shows No in reverse video.
Press the <Measurement key> if you wish to terminate the procedure.
Select Yes by turning the <Control ball>.
Press the <Measurement key> if you wish to set default values.
The display shows Accept data in reverse video.
Press the <Measurement key> to set all jobs to the default values.
5.3.1.3 Reject data
Procedure:
Select the menu line Reject data by turning the <Control ball> and confirm this action by pressing
the <Measurement key>.
The new data will not be stored; stored data are retained.
The menu line Return is shown in reverse video on the display.
Press the <Measurement key> several times to return to the main menu.
Vue de la page 65
1 2 ... 61 62 63 64 65 66 67 68 69 70 71 ... 163 164

Commentaires sur ces manuels

Pas de commentaire